Spectrophotometer-UV-Visible
Evolution 201/220 UV-Visible Spectrophotometers
For routine to research analysis, this instrument series offers real innovations that provide enhanced usability and performance – without added complexity. With straightforward software, cutting-edge instrumentation and a wide selection of accessories, the Evolution 201 and 220 systems work with you to perform experiments the way you want – for the results you need.
INSIGHT Software—Sophistication Simplified
- Personalize your home screen for easy navigation and convenience
- Workbook tools save time and ensure accurate results
- CUE scripting allows creation of customized interface and simplified workflow
Optimized Engineering Delivers Ultimate Performance
- Double-beam geometry with 1nm spectral bandwidth for precision measurements
- Application Focused Beam Geometry (AFBG) technology enables the best performance from microcells, solid samples, and fiber optic probes
- Xenon flash lamp requires no warm-up time allowing instant measurements
- Electronic triggering provides the highest level of accuracy for rapid-mixing kinetics measurements
- Precision monochromator drive delivers fast scanning data collection and wavelength accuracy
Powerful Convenience at your Fingertips
- Color, touch screen provides instrument control using built-in computer
- Integrated keypad allows communication with INSIGHT software
- Choice of Local or PC configurations provides flexibility and convenience
Accessories for all your Sampling Needs
- Wide accessory selection with various temperature control options, sample changers, sippers and solid sampling devices complete your analysis
- Choice of Peltier Single Cell Holder or a Smart 8-Cell Peltier system for temperature control and sample monitoring from 0° to 100 °C
- Easy software integration gives you complete control of your measurements
Assurance of System Performance
- The Thermo Scientific UV Validator package provides support for system qualification and validation activities
- Documentation facilitates compliance of your system with the requirements of FDA,GxP, ISO 9001:2000 and ISPE 2001 guidelines and regulations
- Mercury Lamp accessory is available to perform instrument calibration
- Thermo Scientific INSIGHT security software provides complete tools to meet 21 CFR Part 11 requirements
Specifications |
Model |
Evolution 201 |
Evolution 220 |
Spectral Bandwidth |
1nm |
Variable: 1nm; 2nm
AFBG Microcell optimized
AFBG Fiber optic optimized
AFBG Materials optimized |
Optical Design |
Double Beam with sample and reference cuvet positions; Czerny-Turner Monochromator |
Light Source |
Xenon flash lamp, 3-year warranty (Typical lifetime: 5 years) |
Detector |
Dual Silicon Photodiodes |
Wavelength |
Range |
190 to 1100nm |
Accuracy |
±0.8nm (full range 190 to 1100nm)
± (546.11nm mercury line) |
Scan Speed |
<1 to 6000nm/min.; variable |
Photometric |
Range |
>3.5A |
Accuracy |
0.5A: ±0.0004A
1A: ±0.006A
2A: ±0.010A
Measured at 440nm using neutral density filters traceable to NIST*/NLPL |
Noise |
0A: <=0.00015 A
1A: <=0.00050 A
2A: <=0.00080 A
260nm, 1nm SBW, RMS |
Dimensions |
24.3 x 19.1 x 11 in. (62.2 x 48.6 x 27.9mm) |
Weight |
31.7 lb. (14.4kg) |
Electrical Requirements |
90 to 240VAC 47 – 63Hz, Single Fuse (110V vs. 220V) |
Cat. No. |
Description |
840-210800 |
Evolution 201 PC |
840-210700 |
Evolution 201 LC |
840-210600 |
Evolution 220 PC |
840-210500 |
Evolution 220 LC |
]]>